The role of mass spectrometry in purity verification

Tide Front Supply Research Team · Apr 4, 2026 · 1 min read

Why HPLC purity numbers alone don't prove identity, and how MS data close the loop.

HPLC tells you a sample is clean. Mass spectrometry tells you it's the right thing. Both matter; neither replaces the other.

What MS measures

A peptide's molecular weight is determined by its amino-acid composition. ESI-MS or MALDI-TOF can measure that weight to within a fraction of a Dalton, distinguishing the target sequence from common synthesis byproducts.

Common confounders

  • Truncated sequences: missed couplings during synthesis can produce peptides one or two residues short.
  • Oxidation: methionine and tryptophan oxidize during synthesis or storage, shifting mass by +16 Da per event.
  • Deamidation: asparagine and glutamine convert to aspartate and glutamate over time, shifting mass by +1 Da.

What we report

Every CoA includes the observed mass and the theoretical mass. A delta of less than 1 Da on a peptide under 5 kDa is normal instrument tolerance; a delta of 16 Da means oxidation; a delta of -114 Da means a missing amino acid.

If those numbers don't match what's claimed, don't run the experiment.